Chapter 3 61
Mode
Meas Off (Spectrum Analyzer) Measurement
Mode
Limit Mask 1 (upper) or Limit Mask 2 (Lower)
Limit Mask 1 (upper) access the upper limit mask keys.
Limit Mask 2 (Lower) access the lower limit mask keys.
You can define a limit mask that matches your maximum or minimum measurement
requirements. This mask can then be compared with the measured trace data to determine
pass/fail conditions.
Limit Trace
Selects the trace to which the currently selected limit mask (1 or 2) will be applied. The limit masks
can be applied to separate traces. Only 2 traces are available in the Stimulus Response mode
(compared to 4 traces in the Spectrum Analyzer Mode).
Mode Spectrum Analyzer
Dependencies /
Couplings
The Limit Mask 1 (Upper) menu link key accesses the keys to
configure the upper limit mask for the Center, 1
st
, or 2
nd
segments (dependent on selected segment).
The Limit Mask 2 (Lower) menu link key accesses the keys to
configure the lower limit mask for the Center, 1
st
, or 2
nd
segments (dependent on selected segment).
Remote Command
Notes
The Limit Masks SCPI command syntax uses:
The keyword “LMASk[1]” to define the Upper limit mask.
The keyword “LMASk2” to define the Lower limit mask.
State Saved Saved in instrument state.
Key Path
Meas Setup, Limit Masks
Mode SA
Remote Command :CALCulate:LMASk[1]|2:TRACe 1|2|3|4
:CALCulate:LMASk[1]|2:TRACe?
Example :CALC:LMAS1:TRAC 2
Dependencies /
Couplings
The limit mask is always displayed when the center segment is
turned on; even if the trace to which it is being applied is
blanked (‘Show’ set to Off).
Preset 1
State Saved Saved in instrument state.