74 Chapter 3
Mode
Meas Off (Spectrum Analyzer) Measurement
Mode
Trig Slope Controls the trigger polarity. It is set positive to trigger on a rising edge and
negative to trigger on a falling edge.
Trig Delay Controls a time delay during which the analyzer will wait to begin a sweep
after meeting the trigger criteria. You can use negative delay to pre-trigger the instrument
in time domain, but not in swept spans.
Dependencies / Couplings The range of the Video Trigger Level is dependent on the Reference
Level.
Trigger level selection is not available with free run or external
triggering.
Key Path
Meas Setup, Trigger, Trigger Level
Remote Command :TRIGger[:SEQuence]:EXTernal:SLOPe POSitive|NEGative
:TRIGger[:SEQuence]:EXTernal:SLOPe?
Example TRIG:EXT:SLOP NEG
Preset POSitive
State Saved Saved in State
Dependencies / Couplings Trigger slope selection is only available with external triggering.
Key Path
Meas Setup, Trigger, Trig Slope
Remote Command :TRIGger[:SEQuence]:DELay <time>
:TRIGger[:SEQuence]:DELay?
Example TRIG:DEL 100 ms
Preset 1 µs
State Saved Saved in State
Min -204.8 s
Note this is RBW dependent: -102400 / RBW (Hz) with a maximum
of -27 ms
Max +4.5 hours
Note this is RBW dependent: -102400 / RBW (Hz) with a minimum
of +2 s
Dependencies / Couplings Trigger delay selection is not available with free run triggering.