212 Chapter 3
Mode
Return Loss
Mode
Center Segment
Segments (Center) Sets the state of the center segment to On or Off. This key is available only when Center has
been selected with the Segments key.
Level: Center
Sets the power level for the center segment of the current selected limit mask. This key is available only when
Center has been selected with the Segments key.
Span: Center
Sets the frequency span for the center segment of the currently selected limit mask. This key is only available
when Center has been selected with the Segment key.
Notes Key is only displayed when the selected segment is Center
Dependencies /
Couplings
If Off no limit mask will be displayed even if the 1
st
or 2
nd
segments are turned On.
If On the center segment is drawn on the display centered
round the current center frequency.
Preset OFF
State Saved Saved in instrument state.
Range On | Off
Key Path
Meas Setup, Limit Masks, Limit Mask 1, Center Segment
Mode Stimulus / Response
Restrictions and Notes Key is only displayed when the selected segment is Center.
Preset Not affected by a preset but set per default to 0 dB after a power
cycling
State Saved Saved in instrument state.
Min –10 dB
Max 150 dB
Key Path
Meas Setup, Limit Masks, Limit Mask 1, Center Segment Level
Default Unit dB
Mode Stimulus / Response
Restrictions and Notes Key is only displayed when the selected segment is Center.