264 Chapter 4
Utility and Control/Sweep Functions
Utility Functions
Utility and Control/Sweep Functions
Type
Enables you to select the type of information you want to save. The file types available for
saving are described below.
Remote Command Notes For the :MMEMory:STORe:TRACe <label>,<“file_name”>
command:
Trace names are:
TRACE1|TRACE2|TRACE3|TRACE4|ALL
Measurements in the Stimulus/Response mode have only
TRACE1|TRACE2 available
The file name must have a file extension of .trc.
The <‘file_name’> must include the complete path, for
example ‘C:/MyTrace.trc’
When you are making measurements in the
Stimulus/Response Mode, the :MMEMory:STORe:STATe
1,<‘file_name’> command is only supported in the Two-Port
Insertion Loss measurement. It is not supported in the
One-Port Insertion Loss, Return Loss, or Distance To Fault
measurements.
Key Path
Save, Save Now
Type Format Source Extension
Screen Bit-Mapped Graphic png
Reverse Bit-Mapped
Graphic (light
background)
png
State State sta
Trace Trace + state Trace 1, 2, 3, 4, or
all traces
trc
Measurement Results
Data
Spreadsheet Trace 1 xls
Mode All
Preset Type survives Preset and *RST, but is set to State at power on.
State Saved Type is not saved in the instrument state