318 Chapter 5
Programming Fundamentals
Using the LAN to Control the Instrument
Programming Fundamentals
Check the physical connection to the LAN.
Make sure the internet (IP) Address of the instrument is set up
correctly in the LAN port setup menu. (Press
System, Controls, IP
Admin
.)
When connecting to your instrument over a closed network
(directly through a hub or crossover cable) it may help to set the
instrument to its default settings for subnet mask and gateway.
(subnet mask: 255.255.0.0, gateway 0.0.0.0)
If the instrument and the computer are on different networks or
subnets, make sure the gateway address and subnet mask values
are set correctly.
Cannot access the file system via ftp
If you get a “connection refused” message, try the following
solutions:
If the power to the instrument was just turned on, make sure that
you wait about 25 seconds before attempting the connection.
If you get a “connection timed out” message
Verify the LAN connection between your computer and the
instrument. Refer to “If you cannot connect to the instrument”
earlier in this section.
Cannot telnet to the command parser port
For a “connection refused” message
Check the telnet port number from the front-panel keys.
For a “connection timed out” or “no response from host” message
Verify the LAN connection between your computer and the
instrument. Refer to “If you cannot connect to the instrument”
earlier in this section.
For a “connection refused” or “no response from host” message
If the instrument was just turned on, make sure that you wait
about 25 seconds before attempting the connection.
An “operation timed-out” message
Check the LAN connection between the computer and the
instrument. Refer to “If you cannot connect to the instrument” in
this section.
Increase the file time-out value on your PC or workstation.
Cannot access internal web pages or import graphic images
when using a point-to-point connection