Chapter 6 329
Using The Status Monitoring Subsystem
Using a Status Register
Using The Status Monitoring
Subsystem
5. You can do a similar thing with this register to only look at bit 9 using,
STAT:QUES:ENAB 512.
6. The Status Questionable register output goes to the “Status Questionable Summary”
bit 3 of the Status Byte Register. The output from this register can be enabled using the
*SRE 8 command.
7. Finally, you would use the serial polling functionality available for the particular
bus/software that you are using to monitor the Status Byte Register. (You could also use
*STB? to poll the Status Byte Register.)