Chapter 6 331
Using The Status Monitoring Subsystem
Using the Service Request (SRQ) Method
Using The Status Monitoring
Subsystem
measurement is in continuous mode, then restarting a measurement (INIT command) can
cause the measuring bit to pulse low. This causes an SRQ when you have not actually
reached the “end-of-measurement” condition. To avoid this:
Set INITiate:CONTinuous off.
Set/enable the status registers.
Restart the measurement (send INIT).