Chapter 7 Remote Commands
7-18
:DEMux:DATA:INTerface <input>
Parameter <input>=<CHARACTER PROGRAM DATA>
SINGle Single Ended
DIF50ohm Differential 50
DIF100ohm Differential 100
Function Sets data input interface.
Example To set Data Interface interface to single connection:
> :DEMux:DATA:INTerface SINGle
Compatibility Compatible with MU182040A and MU182041A.
:DEMux:DATA:INTerface?
Response <interface>=<CHARACTER RESPONSE DATA>
SING, DIF50, DIF100
Function Queries data input interface.
Example To query Data Interface interface:
> :DEMux:DATA:INTerface?
< SING
Compatibility Compatible with MU182040A and MU182041A.
:DEMux:DATA:DIFFerential <item>
Parameter <item>=<CHARACTER PROGRAM DATA>
INDependent Data/Xdata independent setting
TRACking Data/Xdata common setting
ALTernate Data/Xdata differential setting
Function Sets the differential type when the data input interface is differential.
Example To set the differential type for Data1Interface to Independent:
> :DEMux:DATA:DIFFerential INDependent
Compatibility Compatible with MU182040A and MU182041A.
:DEMux:DATA:DIFFerential?
Response <item>=<CHARACTER RESPONSE DATA>
IND, TRAC, ALT
Function Queries the differential type when the data input interface is set to
Differential.
Example To query the differential type for Data1Interface:
> :DEMux:DATA:DIFFerential?
< IND
Compatibility Compatible with MU182040A and MU182041A.