7.3 64G DEMUX Commands
7-19
7
Remote Commands
:DEMux:DATA:SINGle <item>
Parameter <item>=<CHARACTER PROGRAM DATA>
DATA Data is selected
XDATa XData is selected
Function Sets input port for Single data input.
Example To set Data Interface input port to data:
> :DEMux:DATA:SINGle DATA
Compatibility Compatible with MU182040A and MU182041A.
:DEMux:DATA:SINGle?
Response <item>=<CHARACTER RESPONSE DATA>
DATA Data is selected
XDAT XData is selected
Function Queries input port for Single data input.
Example To query Data Interface input port:
> :DEMux:DATA:SINGle?
< DATA
Compatibility Compatible with MU182040A and MU182041A.
:DEMux:DATA:THReshold <port>,<numeric>
Parameter <port>=<CHARACTER PROGRAM DATA>
DATA, XDATa (At Single, input the selected port.)
<numeric>=<DECIMAL NUMERIC PROGRAM DATA>
–3.500 to 3.300 –3.500 to 3.300 V/0.001 V Step
Function Sets date input threshold value for specific port.
Example To set threshold value for Data1Interface input port to –0.5 V:
> :DEMux:DATA:THReshold DATA,-0.5
Compatibility Compatible with MU182040A and MU182041A.
:DEMux:DATA:THReshold? <port>
Parameter <port>=<CHARACTER PROGRAM DATA>
DATA, XDATa (At Single, input the selected port.)
Response <numeric>=<NR2 NUMERIC RESPONSE DATA>
Function Queries date input threshold value for specific port.
Example To query threshold value for Data Interface input port:
> :DEMux:DATA:THReshold? DATA
< -0.500
Compatibility Compatible with MU182040A and MU182041A.