7.3 64G DEMUX Commands
7-35
7
Remote Commands
:DEMux:CALCulate:DATA:MONitor? <item>
Parameter <item>=<STRING PROGRAM DATA>
"BIT:TOTal" Bit Error (Total Error)
"PSLoss" Pattern Sync Loss
"CALarm" Clock Alarm
Response <string>=<STRING RESPONSE DATA>
Table 7.3.2.1-4 Response Format
Form Format Description
Form1 "Occur" When an alarm occurs
"Not Occur" When no alarm occurs
"-------" When no data corresponds to the query
Function Queries real-time occurrence status of monitor items specified by
parameters.
Example To query Bit Error occurrence status:
> :DEMux:CALCulate:DATA:MONitor? "BIT:TOTal"
< "Occur"
Compatibility Compatible with MU182040A and MU182041A.