Chapter 7 Remote Commands
7-46
:SENSe:MEASure:STARt
Function Starts the Eye Margin measurement.
Example > :SENSe:MEASure:STARt
Compatibility Incompatible with existing models.
:SENSe:MEASure:STOP
Function Stops the Eye Margin measurement.
Example > :SENSe:MEASure:STOP
Compatibility Incompatible with existing models.
:SENSe:MEASure:EMARgin:STATe?
Response <boolean>=<NR1 NUMERIC RESPONSE DATA>
1 Starts measurement
0 Stops measurement
–1 Measurement failure
Function Queries the Eye Margin measurement state.
Example > :SENSe:MEASure:EMARgin:STATe?
< 1
Compatibility Compatible with the MP1632C.
:SENSe:MEASure:EMARgin:MARGin:THReshold <thre>
Parameter <thre>=<CHARACTER PROGRAM DATA>
E_3 to E_12 1E–3 to 1E–12/1 Step
Function Sets the threshold value that is to be the boundary of the Eye Margin
measurement margin.
Example To set the threshold value that is to be the Eye Margin measurement
margin to 1E–4:
> :SENSe:MEASure:EMARgin:MARGin:THReshold E_4
Compatibility Compatible with the MP1632C.
:SENSe:MEASure:EMARgin:MARGin:THReshold?
Response <thre>=<CHARACTER RESPONSE DATA>
E_3 to E_12 1E–3 to 1E–12
Function Queries the threshold value that is to be the boundary of the Eye Margin
measurement margin.
Example > :SENSe:MEASure:EMARgin:MARGin:THReshold?
< E_4
Compatibility Compatible with the MP1632C.