7.4 Automatic Measurement Commands
7-55
7
Remote Commands
:SENSe:MEASure:EDIagram:DIAGram:SLOT
<slot>,<interface>,<boolean>[,<unit>]
Parameter <slot>=<CHARACTER PROGRAM DATA>
SLOT1 to SLOT6 Slots No. 1 to No. 6
1 to 6 when using the MP1800A, 1 to 4 when using the MT1810A
USB11 to USB14 USB No.11 to 14
<interface>=<DECIMAL NUMERIC PROGRAM DATA>
1 to 4 Data1 to 4
Fixed to 1 if using a USB module<boolean>=<BOOLEAN PROGRAM
DATA>
ON or 1 Eye Diagram measurement ON
OFF or 0 Eye Diagram measurement OFF
[<unit>]=<DECIMAL NUMERIC PROGRAM DATA>
1 to 4 Mainframe Nos. 1 to 4
When using two or more MT1810A units in serial connection, specify the
mainframe number.
Can be omitted. Mainframe No. 1 is specified when omitted.
Function Selects the Eye Diagram measurement slot.
Example To set USB No.11 as a target of Eye Diagram measurement:
> :SENSe:MEASure:EDIagram:DIAGram:SLOT USB11,1,ON
Compatibility Incompatible with existing models.
:SENSe:MEASure:EDIagram:DIAGram:SLOT?
<slot>,<interface>[,<unit>]
Parameter <slot>=<CHARACTER PROGRAM DATA>
SLOT1 to SLOT6 Slots No. 1 to No. 6
1 to 6 when using the MP1800A, 1 to 4 when using the MT1810A
USB11 to USB14 USB No.11 to 14
<interface>=<DECIMAL NUMERIC PROGRAM DATA>
1 to 4 Data1 to 4
Fixed to 1 if using a USB module
[<unit>]=<DECIMAL NUMERIC PROGRAM DATA>
1 to 4 Mainframe Nos. 1 to 4
When using two or more MT1810A units in serial connection, specify the
mainframe number.
Can be omitted. Mainframe No. 1 is specified when omitted.
Response <boolean>=<NR1 NUMERIC RESPONSE DATA>
ON or 1 Eye Diagram measurement ON
OFF or 0 Eye Diagram measurement OFF
Function Queries whether the specified slot is targeted for the Eye Diagram
measurement.
Example To query whether USB No.11 is subject to measurement:
> :SENSe:MEASure:EDIagram:DIAGram:SLOT? USB11,1