7.4 Automatic Measurement Commands
7-61
7
Remote Commands
:DISPlay:RESult:EDIagram:SLOT <slot>,<interface>[,<unit>]
Parameter <slot>=<CHARACTER PROGRAM DATA>
SLOT1 to SLOT6 Slots No. 1 to No. 6
1 to 6 when using the MP1800A, 1 to 4 when using the MT1810A
USB11 to USB14 USB No.11 to 14
<interface>=<DECIMAL NUMERIC PROGRAM DATA>
1 to 4 Data1 to 4
Fixed to 1 if using a USB module
[<unit>]=<DECIMAL NUMERIC PROGRAM DATA>
1 to 4 Mainframe Nos. 1 to 4
When using two or more MT1810A units in serial connection, specify the
mainframe number.
Can be omitted. Mainframe No. 1 is specified when omitted.
Function Selects the Eye Diagram result display target slot.
Example To set the Eye Diagram result display target slot to slot No. 1, Data 1:
> :DISPlay:RESult:EDIagram:SLOT SLOT1,1
Compatibility Incompatible with existing models.
:DISPlay:RESult:EDIagram:SLOT? [<unit>]
Parameter [<unit>]=<DECIMAL NUMERIC PROGRAM DATA>
1 to 4 Mainframe Nos. 1 to 4
When using two or more MT1810A units in serial connection, specify the
mainframe number.
Can be omitted. Mainframe No. 1 is specified when omitted.
USB11 to USB14 USB No.11 to 14
Fixed to 1 if using a USB module
Response <slot>=<CHARACTER RESPONSE DATA>
SLOT1 to SLOT6 Slots No. 1 to No. 6
<interface>=<NR1 NUMERIC PROGRAM DATA>
1 to 4 Data1 to 4
Function Queries the Eye Diagram result display target slot.
Example > :DISPlay:RESult:EDIagram:SLOT?
< SLOT1,1
Compatibility Incompatible with existing models.