EasyManua.ls Logo

Anritsu MT1810A - Page 181

Anritsu MT1810A
272 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7.4 Automatic Measurement Commands
7-63
7
Remote Commands
:DISPlay:RESult:EDIagram:SCALe:VOLTage <max>,<step>
Parameter <max>=<DECIMAL NUMERIC PROGRAM DATA>
–3.990 to 4.000 –3.990 to 4.000 V/0.001 V Step
<step>=<DECIMAL NUMERIC PROGRAM DATA>
0.001 to 0.800 0.001 to 0.800 V/0.001 V Step
The range varies depending on the option status and input interface setting.
Function Sets the maximum value in the direction of the Threshold of the Eye
Diagram graph, and the grid resolution.
Example To set the maximum value in the direction of the Threshold of the Eye
Diagram graph to 4.000 V, and the grid resolution to 0.800 V:
> :DISPlay:RESult:EDIagram:SCALe:VOLTage 4.000,0.800
Compatibility Incompatible with existing models.
:DISPlay:RESult:EDIagram:SCALe:VOLTage?
Response <max>=<NR2 NUMERIC RESPONSE DATA>
–3.990 to 4.000 –3.990 to 4.000 V
<step>=<NR2 NUMERIC RESPONSE DATA>
0.001 to 0.800 0.001 to 0.800 V
Function Queries the maximum value in the direction of the Threshold of the Eye
Diagram graph, and the grid resolution.
Example > :DISPlay:RESult:EDIagram:SCALe:VOLTage?
< 4.000,0.800
Compatibility Incompatible with existing models.
:DISPlay:RESult:EDIagram:SCALe:PHASe <max>,<step>
Parameter <max>=<DECIMAL NUMERIC PROGRAM DATA>
–990 to 1000 –990 to 1000 mUI/1 mUI Step
<step>=<DECIMAL NUMERIC PROGRAM DATA>
1 to 200 1 to 200 mUI/1 mUI Step
Function Sets the maximum value in the direction of the Phase of the Eye Diagram
graph, and the grid resolution.
Example To set the maximum value in the direction of the Phase of the Eye Diagram
graph to 1000, and the grid resolution to 10 mUI:
> :DISPlay:RESult:EDIagram:SCALe:PHASe 1000,10
Compatibility Incompatible with existing models.

Table of Contents

Other manuals for Anritsu MT1810A

Related product manuals