Chapter 7 Remote Commands
7-80
:DISPlay:RESult:EDIagram:TEMPlate:PATTern <type>
Parameter <type>=<CHARACTER PROGRAM DATA>
E1 Edit1
E2 Edit2
E3 Edit3
E4 Edit4
Function Sets the pattern to be edited during Eye Diagram measurement.
Example To set the pattern to be edited during Eye Diagram measurement to Edit4:
> :DISPlay:RESult:EDIagram:TEMPlate:PATTern E4
Compatibility Incompatible with existing models.
:DISPlay:RESult:EDIagram:TEMPlate:PATTern?
Response <type>=<CHARACTER RESPONSE DATA>
E1, E2, E3, E4
Function Queries the pattern that was edited during Eye Diagram measurement.
Example > :DISPlay:RESult:EDIagram:TEMPlate:PATTern?
< E4
Compatibility Incompatible with existing models.
:DISPlay:RESult:EDIagram:TEMPlate:ASCale
Function Executes automatic adjustment of the Mask Edit measurement graph scale.
Example > :DISPlay:RESult:EDIagram:TEMPlate:ASCale
Compatibility Incompatible with existing models.
:DISPlay:RESult:EDIagram:TEMPlate:SCALe <max>,<step>
Parameter <max>=<DECIMAL NUMERIC PROGRAM DATA>
–3.990 to 4.000 –3.990 to 4.000 V/0.001 V Step
<step>=<DECIMAL NUMERIC PROGRAM DATA>
1 to 800 1 to 800 mV/1 mV Step
Function Sets the maximum value and the grid resolution in the direction of the
Threshold of the Mask Edit graph.
Example To set the maximum value and the grid resolution in the direction of the
Threshold of the Mask Edit graph to 4.000 V and 800 mV, respectively:
>:DISPlay:RESult:EDIagram:TEMPlate:SCALe 4.000,800
Compatibility Incompatible with existing models.