Chapter 7 Remote Commands
7-82
:DISPlay:RESult:EDIagram:TEMPlate:PNUMber?
Response <numeric>=<NR1 NUMERIC RESPONSE DATA>
0 to 32 0 to 32 points
Function Queries the number of Mask points.
Example > :DISPlay:RESult:EDIagram:TEMPlate:PNUMber?
< 0
Compatibility Incompatible with existing models.
:DISPlay:RESult:EDIagram:TEMPlate:EDIT <point>,<phase>,<voltage>
Parameter <point>=<DECIMAL NUMERIC PROGRAM DATA>
1 to 32 1 to 32 points, in 1-point steps
<phase>=<DECIMAL NUMERIC PROGRAM DATA>
–1000 to 1000 –1000 to 1000 mUI/1 mUI Step
<voltage>=<DECIMAL NUMERIC PROGRAM DATA>
–4.000 to 4.000 –4.000 to 4.000 V/0.001 V Step
Function Sets a Mask point position.
Example To set a Mask point position (1 point, 1000 mUI, 4.000 V):
> :DISPlay:RESult:EDIagram:TEMPlate:EDIT 1,1000,4.000
Compatibility Incompatible with existing models.
:DISPlay:RESult:EDIagram:TEMPlate:EDIT? <point>
Parameter <point>=<DECIMAL NUMERIC PROGRAM DATA>
1 to 32 1 to 32 points
Response <phase>=<NR1 NUMERIC RESPONSE DATA>
<voltage>=<NR1 NUMERIC RESPONSE DATA>
Function Queries a Mask point position.
Example To query the position of point 1:
> :DISPlay:RESult:EDIagram:TEMPlate:EDIT? 1
< 1000,4.000
Compatibility Incompatible with existing models.