7.4 Automatic Measurement Commands
7-95
7
Remote Commands
> :SENSe:MEASure:BATHtub:ECOunt FINE
Compatibility Incompatible with existing models.
:SENSe:MEASure:BATHtub:ECOunt?
Response <numeric>=<CHARACTER RESPONSE DATA>
FINE, COAR
Function Queries the minimum error counts of Bathtub measurement.
Example > :SENSe:MEASure:BATHtub:ECOunt?
< FINE
Compatibility Incompatible with existing models.
:SENSe:MEASure:BATHtub:RANGe <range>
Parameter <range>=<CHARACTER PROGRAM DATA>
E_7 to E_14 E–7 to E–14
Function Sets the lower limit value of measurement error threshold during Bathtub
measurement.
Example To set the lower limit value of measurement error threshold to E–12:
> :SENSe:MEASure:BATHtub:RANGe E_12
Compatibility Incompatible with existing models.
:SENSe:MEASure:BATHtub:RANGe?
Response <range>=<CHARACTER RESPONSE DATA>
Function Queries the lower limit value of measurement error threshold during
Bathtub measurement.
Example > :SENSe:MEASure:BATHtub:RANGe?
< E_12
Compatibility Incompatible with existing models.
:SENSe:MEASure:BATHtub:CALRange <upper>,<lower>
Parameter <upper>=<CHARACTER PROGRAM DATA>
E_3 to E_5 E–3 to E–5
<lower>=<CHARACTER PROGRAM DATA>
E_7 to E_14 E–7 to E–14
Function Sets the error threshold range during Bathtub measurement recalculation.
Example To set the error threshold range during recalculation to E–5 to E–10:
> :SENSe:MEASure:BATHtub:CALRange E_5,E_10
Compatibility Incompatible with existing models.