Chapter 7 Remote Commands
7-94
:SENSe:MEASure:BATHtub:DATA <numeric>
Parameter <numeric>=<DECIMAL NUMERIC PROGRAM DATA>
–3.500 to 3.300 –3.500 to 3.300 V/0.001 V Step
The range varies according to the option and input interface settings.
Function Sets the measurement voltage threshold position during Bathtub
measurement.
Example To set the measurement voltage threshold position to 0.5 V:
> :SENSe:MEASure:BATHtub:DATA 0.5
Compatibility Incompatible with existing models.
:SENSe:MEASure:BATHtub:DATA?
Response <nuemric>=<NR2 NUMERIC RESPONSE DATA>
Function Queries the measurement voltage threshold position during Bathtub
measurement.
Example > :SENSe:MEASure:BATHtub:DATA?
< 0.505
Compatibility Incompatible with existing models.
:SENSe:MEASure:BATHtub:RESolution:MUI <numeric>
Parameter <numeric>=<DECIMAL NUMERIC PROGRAM DATA>
4 to 100 4 to 100 mUI/4 mUI Step
Function Sets the Bathtub measurement phase resolution (mUI).
Example To set the phase resolution to 10 mUI:
> :SENSe:MEASure:BATHtub:RESolution:MUI 10
Compatibility Incompatible with existing models.
:SENSe:MEASure:BATHtub:RESolution:MUI?
Response <numeric>=<NR1 NUMERIC RESPONSE DATA>
Function Queries the Bathtub measurement phase resolution (mUI).
Example > :SENSe:MEASure:BATHtub:RESolution:MUI?
< 4
Compatibility Incompatible with existing models.
:SENSe:MEASure:BATHtub:ECOunt <mode>
Parameter <mode>=<CHARACTER PROGRAM DATA>
FINE Fine (Number of error counts: 100)
COARse Coarse (Number of error counts: 1)
Function Sets the minimum error counts of Bathtub measurement.
Example To set the minimum error counts to 100: