3.2 Inter-Module Connection
3-13
3
Panel Layout and Connectors
8.
Connect the Delayed Clock Output and DEMUX Clock Input
connectors on the rear panel of MP1862A by using the J1654A cable
that comes with MP1862A.
9.
Connect the Data Output1/2 connectors on the rear panel of
MP1862A and the Data Input1/2 connectors of MU183040B
respectively by using coaxial cables. The coaxial cables must be one
of the following:
• J1657A coaxial cables that come with MP1862A
• Coaxial cables that are of the same length
10.
Connect the 1/2 Clock Output connector on the rear panel of
MP1862A and the Ext. Clock Input connector of MU183040B by
using the J1668A coaxial cable that comes with MP1862A.
11.
Connect the Reference Clock Output connector of MU181500B and
the Ext. Clock Input connector on the front panel of MP1862A by
using a coaxial cable.
Then, on the Interface tab of MP1862A, in the Clock Input Band box,
click the option that matches both bit rate and clock rate to be used
by MP1862A. For details on how to set the Clock Input Band, refer to
4.4.1 “Input setting items”. For bathtub jitter measurement, refer to
4.9 “Bathtub Function”.