C-5 Vector Network Analyzer Specific Warning Messages Error Messages
C-6 PN: 10580-00301 Rev. D MS20xxB UG
PMON PLD Fail
Unable to communicate with the Power Monitor PCBA.
Power Supply
Power Supply failed. Charge the battery.
Error Saving File. General Error Saving File
An error was detected while saving a file. Try again.
C-5 Vector Network Analyzer Specific Warning Messages
Bias Tee cannot be enabled for start freq < 2MHz.
Adjust frequency before turning On the Bias Tee.
The start frequency cannot be set less than 2 MHz when the internal or external Bias Tee is
turned On. Set the frequency to a value larger than or equal to 2 MHz, and then turn on the
Bias Tee.
Bias Tee is not allowed for start freq < 2MHz.
Turn Off Bias Tee before changing the freq.
The internal or external Bias Tee cannot be turned on when the start frequency is set to less
than 2 MHz. Turn off the Bias Tee, and then adjust the start frequency to the desired value
less than 2 MHz.
Changing Source Power
Changing Source Power will affect the accuracy of the current calibration.
While Cal Correction is turned On, changing the source power level will affect the accuracy of
the current calibration. The correction will remain On and can be used, but an indicator (?P)
will appear next to the left-hand status column (CAL: ON) to note that the current power
setting is different from the one used during the calibration processes.
No valid calibration to change correction.
There is no valid calibration in volatile memory that can be used to turn Cal Correction On. A
new calibration must be performed.
Cannot continue with calculating.
Cannot continue with calculating. Not all required cal steps are completed.
When performing a calibration, you must complete all of the required steps before applying
the “Calculate and Finish Cal” step.