Quick Start Guide 3-5 Spectrum Analyzer
MS20xxB UG PN: 10580-00301 Rev. D 3-9
Cal Type
The Cal Type soft key is found in the Calibration Menu (shown in the VNA Measurement
Guide). The Calibration Type list box provides the complete selection of available calibration
types.
3-5 Spectrum Analyzer
Set the instrument to Spectrum Analyzer mode as described in Section 3-2 “Measurement
Mode Selection” on page 3-1.
Set Start and Stop Frequencies
1. Press the Freq main menu key (hard key).
2. Press the Start Freq soft key (submenu key).
3. Enter the desired start frequency using the keypad, the arrow keys, or the rotary knob.
When entering a frequency using the keypad, the soft key labels change to GHz, MHz,
kHz, and Hz. Press the appropriate unit key. Pressing the Enter key has the same affect
as pressing the MHz soft key.
4. Press the Stop Freq soft key.
5. Enter the desired stop frequency.
Enter the Center Frequency
1. Press the Freq main menu key (hard key).
2. Press the Center Freq soft key.
3. Enter the desired center frequency using the keypad, the arrow keys, or the rotary knob.
When entering a frequency using the keypad, the soft key labels change to GHz, MHz,
kHz, and Hz. Press the appropriate unit key. Pressing the Enter key has the same affect
as pressing the MHz soft key.
The center frequency and span is shown at the bottom of the screen.
Select a Signal Standard
1. Press the Freq main menu key (hard key).
2. Press the Signal Standard soft key. The Signal Standards dialog box opens.
3. Highlight a signal standard and press Enter to select.
4. Press the Channel soft key to change the channel value in the Channel Editor.
The signal standard is shown in yellow at the top of the screen.
Set the Measurement Bandwidth
1. Press the BW main menu key (hard key) to display the BW menu.
Press the RBW and/or the VBW soft key to manually change the values.
Set RBW and VBW automatically by pressing the Auto RBW soft key or the
Auto VBW soft key.

Table of Contents

Other manuals for Anritsu VNA Master MS2024B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Anritsu VNA Master MS2024B and is the answer not in the manual?

Anritsu VNA Master MS2024B Specifications

General IconGeneral
BrandAnritsu
ModelVNA Master MS2024B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals