Maintaining the Compounder
Operator Manual for the Baxter ExactaMix 2400 Compounder
Page 102 53007410 Rev. C
1. Shut down and turn off the compounder. Refer to Rebooting and Shutting Down on
Page 32.
2. If the tube set is installed, remove and discard it. Refer to Removing the Expired Tube
Set and Expired Ingredients on Page 44.
3. Open the pump door.
4. Remove and retain the thumbscrew and washer used to attach the rotor.
Removing the thumbscrew
5. Remove the pump rotor from the spindle.
Removing the pump rotor
6. Using the approved materials, clean the:
a. Pump rotor, making sure that the rollers spin freely
b. Pump rotor area
c. Channels near the pump rotor area
7. Install the:
a. Pump rotor, aligning the notch on the bottom of the rotor with the pin on the
spindle
b. Washer and thumbscrew
8. Close the pump door.
9. Using the approved materials, clean the:
a. Valve actuators
b. Surface of the main module
c. Load cell
d. Poles and holders on the vial rack