Setting Up the Compounder
Operator Manual for the Baxter ExactaMix 2400 Compounder
53007410 Rev. C Page 45
Installing the New Valve Set
CAUTION
If the valve set is not installed properly, the compounder cannot be calibrated
accurately.
1. Check that the valve actuators are not broken or damaged.
Valve actuators
Normal
Damaged
Broken
WARNING
Do not use the compounder if a valve actuator is broken or damaged.
Patient harm can result. For assistance, contact Baxter Technical Services.
Refer to Getting Help on Page 22
.
2. Remove the valve set from the packaging.
NOTE: Always use aseptic technique when installing the valve set.
IMPORTANT! Check the expiration date on the valve set before installing it. Do not use a
valve set that is past its expiration date.
3. Place the valve set onto the valve actuators.
4. Gently push the end tabs down and out until you hear a click on each end.
Installing the valve set
NOTE: The appearance of the valve set may differ from the example shown above.
5. Make sure that the valve set is installed securely by pulling up on both ends gently.
CAUTION
Once the valve set has been installed, do not attempt to remove it during
operation.
6. Open the pump door.