Operator Manual for the Baxter ExactaMix 2400 Compounder
53007410 Rev. C Page 203
On-screen Text
Explanation
Suggested Actions
The bag currently on the
load cell does not appear
to be empty.
Before pumping starts, the
load cell detects that the
destination bag contains
fluid.
If the bag is not empty, and you are:
Compounding the solution, refer to Step 2 on
Page 87
Calibrating the compounder, refer to Step 4 on
Page 70
If the bag is empty:
1. Tap No.
2. At the Operation Cancelled message, tap OK.
3. Remove the bag from the load cell. Refer to:
Removing the Patient Bag on Page 90
Removing the Calibration Bag on Page 73
4. Calibrate the load cell. Refer to Calibrating the
Load Cell on Page 37.
5. If necessary, reattach the appropriate bag to
the load cell. Refer to:
Attaching the Patient Bag on Page 85
Attaching the Calibration Bag on Page 73
There does not appear to
be a bag hung on the
scale.
The load cell detects that
the destination bag is not
attached.
If the bag is not attached, attach the appropriate
bag to the load cell. Refer to:
Attaching the Patient Bag on Page 85
Attaching the Calibration Bag on Page 73
If the bag is attached:
1. Tap No.
2. Calibrate the load cell. Refer to
Calibrating the
Load Cell on Page 37.
Issues with the Pump
On-screen Text
Explanation
Suggested Actions
Pump was paused.
The pump was paused
during normal operation.
If the pump was paused by tapping Pause, tap
Resume to continue compounding.
[Error: 01-13-002] Unable
to start the pump
because the pump is in a
fault state.
The pump fault occurred.
1. At the pump screen, tap Stop and follow the
on-screen instructions.
2. Reboot the compounder. Refer to Rebooting
and Shutting Down on Page 32.
Pump faulted. Unable to
close valve. Valve is
moving.
A system fault or power
loss occurred.
1. Write a large “X” on the label of the patient
bag, then remove and discard the bag.
2. Reboot the compounder. Refer to Rebooting
and Shutting Down on Page 32.
Pump is in fault state and
must be reset before use.
Reset the pump?
A pump fault occurred.
Tap Yes to reset the pump.