Setting Up the Options
Operator Manual for the Baxter ExactaMix 2400 Compounder
Page 120 53007410 Rev. C
Password Expiration
Select Use password expiration to place an expiration date on each password, if your facility’s
protocol requires that user passwords must be changed on a regular basis.
NOTE: By default Use password expiration checkbox is checked.
For Days password valid, enter the number of days after which the password expires.
Tip! Baxter recommends entering 90 (or the number specified by your facility’s protocol) for the
Days password valid.
Order Entry Serial Number
For Serial number length, enter the number of maximum length of serial number for a formula
file.
NOTE: If the Order Entry Serial Number is more than 36 characters, part of it may be truncated
in the reports.
Barcode Reader
Select Enable barcode reader to allow the use of a barcode reader for scanning labels during
verification of the setup and for loading formulas. If you select this check box, these options
become available:
• Select Use barcode verification if you want to require the use of a barcode reader for
scanning labels during verification of the setup.
• Select Require barcode to initiate compounding if you want to require the use of a
barcode reader for loading formulas. If this box is selected, the user cannot manually
select a saved formula.
WARNING
It is important to use a barcode reader for scanning labels during verification of the
setup and for loading formulas.
NOTE: Compounder supports both 1D & 2D GS1 barcode formats. However to enable 2D
barcode reading a 2D barcode reader (part number # 6500-0500) is needed.
Default User
If your facility does not require users to log in, select a Default User who is logged in
automatically at startup.
Tip! Baxter does not recommend using this feature.
Cosignature
Select Required for Configuration Verification if you want to require a second user to log in
and verify the configuration. If this option is not selected, the same user can set up and verify
the configuration.
Tip! Baxter strongly recommends requiring a cosignature.