Glossary
Operator Manual for the Baxter ExactaMix 2400 Compounder
53007410 Rev. C Page 201
Issues with the Calibration
Load cell not calibrated.
Must calibrate to
calibrated.
2. Calibrate the load cell. Refer to Calibrating the
Calibration out
of Range
The calibration of the load
cell is out of range.
• The load cell is level and locked into place.
• There is nothing (for example, the outlet
tube, bag or cable for the load cell) touching
the pan or base of the load cell.
• There are no environmental factors (for
example, fans) interfering with the load cell.
2. Calibrate the load cell. Refer to
Calibrating the
Load Cell on Page 37. Read the on-screen
messages carefully, and make sure that you do
not place the calibration weight on the load cell
calibrated since the last
tube set change. This
operation must be
completed prior to
pumping a solution.
Would you like to
calibrated.
2. Calibrate the pump. Refer to Calibrating the
Compounder on Page 69.
pump failed.
1. Check that the valve set is installed properly.
Refer to Installing the New Valve Set on
Page 45.
2. For the source container of water, check the
following conditions. Refer to
Attaching the
New Ingredients and Inlets on Page 48.
• The correct inlet is used.
• The inlet is not kinked.
• The bag is spiked properly. Refer to Page 52.
3. Clean any spills near the pump rotor. Refer to
CLEANING THE COMPOUNDER on Page 101.
4. Calibrate the load cell. Refer to Calibrating the
Load Cell on Page 37.
5. Calibrate the pump. Refer to Calibrating the
You have exceeded the
maximum allowed
calibration limit of
20000mL, the system will
recalibrate now. Do you
prevent auto-calibration
from happening.
Therefore a manual
calibration is required.
2. Calibrate the pump. Refer to Calibrating the
Compounder on Page 69