Using Reports
Operator Manual for the Baxter ExactaMix 2400 Compounder
Page 164 53007410 Rev. C
Sample MixCheck Report (right side)
Verify that the Date/Time matches the date and time
the order was fulfilled.
Compare the Measured Weight of the solution to the
Expected Weight. Verify that the Difference is within
acceptable limits per your facility’s protocol.
Verify that each ingredient and its Ordered Volume
match the label on the patient bag and/or the original
formula.
Verify that the Total (mL) matches the label on the
patient bag and/or the original formula.
NOTE: The Total (mL) doesn’t include the volume of
the Manual Additions.