Glossary
Operator Manual for the Baxter ExactaMix 2400 Compounder
Page 194 53007410 Rev. C
Unable to retrieve
formula for scanned
barcode <Formula File
Name> from the formula
file directory <formula file
directory>. There may be
an issue with network
connectivity, please
reboot the compounder.
If the problem persists,
please contact Baxter
The .PAT /.FRM file is not
available.
1. Correct the .PAT/.FRM file in the order entry
software, then print and scan a new barcode
label.
2. Check that the Ethernet cable is connected
properly to both the display and the order-
entry computer.
3. Reboot the compounder. Refer to
Rebooting
and Shutting Down on Page 32.
4. Check that the network is functioning.
5. Check the path on both the order-entry
computer and the compounder. Refer to
Setting Up the Directories Options
on Page 121.
2D Formula Barcode serial
number (X) does not
meet current Order Entry
specification and cannot
serial number does not
meet the Order entry
specification
1. Correct the serial number of the 2D Formula
Barcode, then print and scan the new 2D
Formula Barcode.
The following formula
name or data is not
supported.
Contact the Order Entry
System Administrator.
does not meet the Order
entry specification
1. Correct the content of the 2D Formula Barcode,
then print and scan the new 2D Formula
Barcode.
2D Formula Barcode serial
number format (X) does
not meet the current
configured maximum
serial number length.
Refer to security tab in
options screen or contact
the Order Entry systems.
2D Formula Barcode serial
number format (X) does
not meet the configured
maximum serial number
length, set in options
screen security tab.
The serial number length
should be less than or
equal to the value
specified in the security
1. Either, correct the length of the serial number
of the 2D Formula Barcode (Maximum up to
the value specified in the security tab), then
print and scan the new 2D Formula Barcode
(or)
2. Increase the serial number length value in the
options screensecurity tab, such that the
serial number length available in the 2D
Formula Barcode is lesser than or equal to the
configured value in security tab.