Setting Up the Compounder
Operator Manual for the Baxter ExactaMix 2400 Compounder
53007410 Rev. C Page 55
o To spike and hang a syringe:
a. Turn the syringe with its Luer end facing down.
NOTE: This step helps to reduce the possibility
of air bubbles entering the inlet tube.
NOTE: Only syringes that have been verified by
a pharmacist should be hung on the
compounder.
b. Rotate the inlet onto the syringe.
Rotating
c. Hang the syringe on the vial rack by snapping
the syringe flanges into the syringe holder.
Hanging
6. Label the inlet with the appropriate, numbered inlet barcode label that was packaged
with the valve set. Attach the label close to the source container. The number on the
label must match the number of the port to which the inlet is attached.
7. Check that the correct inlet is:
a. Attached to the correct ingredient and port
b. Labeled with the correct barcode label
Tip! Baxter recommends rotating the source container so that the barcode label faces
you, for easy scanning during barcode verification.
8. At the ingredient detail window, tap OK.