Usage specifications
12
7990087_030_03 – 2080492 – 2023-01-19
1.6 Improper use
– The light suspension unit must not be exposed to additional
loads.
– The surgical light may not be exposed to strong vibrations.
– The surgical light system must not be used for investigatory
and diagnostic purposes.
1.7 User definition
The users must be authorized for the use and service of the
product. Users can be sterile and non-sterile employees in the
operating room, such as surgeons, nurses, cleaning personnel, as
well as service, maintenance and hospital technicians.
1.8 Usage environment
For integrated systems with RS232 interface, a maximum
operating altitude of 2000 m / 6562 ft above sea level applies.
1.9 Ambient conditions for storage and transport
Temperature: +10°C to +35°C / +50°F to +95°F
Air humidity: 30% to 75%
Atmospheric pressure: 70 kPa to 106 kPa / 10 psi to 15 psi
Operating altitude: Up to 3,000 m / 9,843 ft above sea level
Temperature: -20°C to +40°C / -4°F to +104°F
A maximum storage temperature of
+60°C / +140°F applies for the lamp head.
Air humidity: 5% to 95%
Atmospheric pressure: 50 kPa to 106 kPa / 7 psi to 15 psi