EasyManuals Logo

baxter iLED 7 User Manual

baxter iLED 7
90 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #41 background imageLoading...
Page #41 background image
Use
7990087_030_03 – 2080492 – 2023-01-19 41
10. Check the sterilizable handle for visible damage and a secure
fit on the lamp head.
11. Check the functionality of the surgical light with all control
modules in use.
Immediately take a faulty surgical light or control module out of
service, label it clearly as being defective and ensure it is not used
again. If there is damage or a fault, notify the technical customer
service.
Following any damage reported by a user, a functional check and
a visual inspection must be carried out by personnel trained to
perform these tasks by Baxter.
Weekly inspection At least once a week, a functional check and a visual inspection
must be carried out by personnel trained to perform these tasks
by Baxter.
If there is damage or a fault, notify Technical Customer Service.
Annual inspection The annual inspection must be performed by staff trained in this
work by Baxter.
The annual inspection includes a visual inspection and functional
inspection of the entire surgical light considering the following
inspection points.
Check the components of the lamp head and the support arm
system for deformation
Surface damage
Check the support arm system and lamp head for paint
damage
Check plastic parts for cracks, brittle spots and clouding
–Check welds for cracks
The surgical light must be checked for completeness of all
plastic parts (such as covers and plugs)
Check to ensure that device labels and information labels are
legible.
Check the LEDs for functionality
Check the joints for functionality
Check the position of the end stops
Check the effect of the spring and brake forces
Check the safe mode function
5.3 Selection of functions
The functions of the surgical light are set with the following control
units:
Control element
SLC handle adapter (Sterile Light Control)
Mobile Control 7,9
Wall Control Panel
Operating interface of an external control module
(OP integration systems using RS232 interface)
The commands of the individual control modules are executed in
the following sequence:
1. Control element
2. SLC handle adapter (Sterile Light Control)

Table of Contents