Maintenance
7990087_030_03 – 2080492 – 2023-01-19 73
8 Maintenance
WARNING
Do not carry out any maintenance work during surgery or while
the device is in use.
The surgical light system must be serviced at least every 2 years
after handover to the user. After ten years of operation,
maintenance of the surgical light must be carried out annually.
Product maintenance or the exchange of components may only
be carried out by qualified service technicians.
The contact details of service technicians can be obtained from
the Technical Customer Service at Baxter.
Baxter recommends concluding a maintenance agreement, so
that maintenance can be carried out promptly and reliably.
The automated distance
measurement does not work
(ALC Plus indicator flashes).
Failure of the sensor or
repeated incorrect
measurements.
Deactivate ALC Plus
function with the help of the
Mobile Control 7,9, or
Deactivate ALC Plus
function with the help of the
secure mode, or
Disconnect the power to the
surgical light system to force
a restart.
Different color temperatures in the
light field
The color temperature is only
set on one lamp head.
Switch on the synchronization
function.
Mobile Control 7,9 operation
The surgical light cannot be
operated with the
Mobile Control 7,9.
Fault in the WLAN connection Control lighting intensity using
the controls on the lamp head.
The Mobile Control 7,9 is faulty.
Error Possible cause Correction