Use
7990087_030_03 – 2080492 – 2023-01-19 51
Switching individual surgical lights off:
In the action bar [A] of the TruRemote software, slide the slide
control [B] to the left, from I to 0.
The operating range shows the surgical light and its functions in
gray. The settings areas are hidden. The information area is empty.
Switching all surgical lights off:
1. On the action bar [A], touch the icon [D] to switch all surgical
lights in the lighting system off.
A feedback window is displayed.
2. In the feedback window, touch the OK button to switch off
the surgical light.
All surgical lights in the lighting system are switched off. The
operating range shows the surgical light and its functions in
gray. The settings areas are hidden. The information area is
empty. The icon [D] has a white glowing background.
If the lighting system is not to be switched off, touch the
Cancel button. The feedback window is closed and the
control screen is active once more.
5.11.4 Operating the surgical light
1. Switch on the Mobile Control 7,9.
2. Activate the Mobile Control 7,9 / the Wall Control Panel and
unlock it.
3. Mount a sterile handle on each surgical light.
4. Switch on all surgical lights of the lighting system at the
relevant controls or in the action bar of the TruRemote
software.
5. Position all surgical lights of the lighting system.
6. The handle adapter and, where available, the camera of the
lamp heads should be pointed directly at the patient’s wound
area.
7. On the quick-access bar, touch the icon for the desired
surgical light or the icon of the TruVidia Wireless Camera.
The desired surgical light control screen or the
TruVidia Wireless control screen will be shown.
8. In the functions area, touch the icon of the function to be set.
In the settings area of the function, the settings options are
displayed.
9. In the settings area of the function, touch the icon of the
desired setting option.
In the diagram of the surgical light, the change to the setting
is visually displayed.
10. On the action bar, touch the function that is to be carried out
for all surgical lights of the lighting system.
11. Switch the Mobile Control 7,9 / the Wall Control Panel to
standby mode.
12. After surgery, switch off all surgical lights of the lighting
system at the relevant controls or in the action bar of the
TruRemote software.
13. Switch off the Mobile Control 7,9.
14. Connect the Mobile Control 7,9 to the charger for charging.