Cleaning and disinfection
7990087_030_03 – 2080492 – 2023-01-19 69
NOTICE
The patient may be at risk or the product may be damaged when
using unsuitable cleaning agents or disinfectants.
Non-compliance with cleaning or disinfection specifications voids
all warranty claims. No liability is assumed for damage caused by
inappropriate cleaning/disinfection agents. The warranty is valid
only for undamaged surfaces.
Safe use of the surgical light system, the Mobile Control 7,9 and
the Wall Control Panel requires regular cleaning and disinfection
of these components with suitable cleaning agents or
disinfectants after each surgical intervention.
The operator must meet the requirements of the national
Commission for Hygiene and Disinfection.
The cleaning and disinfection of the surgical light system may only
be performed by a hygiene specialist or a person instructed by the
hygiene specialist.
Only agents or chemicals compatibility-tested and approved by
Baxter may be used for cleaning and disinfecting. When using
alternative cleaning agents and disinfectants, Baxter cannot
confirm any material compatibility. Do not use agents that are not
listed, as otherwise functional components may be altered or
damaged.
Before actual disinfection, thorough cleaning of any visible
impurities, such as body fluids, must be performed.
To clean heavy or stubborn dirt, use only a soft brush and mild
cleaner or cleaning disinfectant. Disinfection can take place once
all the visible contaminants have been removed.
After a risk of contamination of the product through potentially
infectious material (e.g. blood, secretions or excrements) disinfect
the surfaces immediately, especially where such material has built
up.
Use only wipe-down disinfection as the disinfection method.
Disinfection using UV light or steam is prohibited.
It is recommended to use the Mobile Control 7,9 after single wipe
disinfection in a protective enclosure. The most suitable product is
Catalyst Case iPad Mini 5 made by Catalyst, with part
number 2074252.
The handles are sterilizable. Take note of the operating
instructions for the handle.