Technical data
82
7990087_030_03 – 2080492 – 2023-01-19
Do not stack the iLED 7 surgical light/Mobile Control 7,9/Wall Control Panel on top of or next to other
devices.
• The iLED 7 surgical light system/Mobile Control 7,9/Wall Control Panel may not be stacked
immediately next to or on top of other devices.
Guidelines and manufacturer's declaration – electromagnetic immunity
Avoid environments with electromagnetic fields and interference stronger than those listed below.
Guidelines and manufacturer's declaration – electromagnetic emissions
The iLED 7 surgical light system/Mobile Control 7,9/Wall Control Panel is intended for use in the
electromagnetic environment indicated below. The customer or user of the iLED 7 surgical light
system/Mobile Control 7,9/Wall Control Panel must ensure that it is operated in such an environment.
Emissions measurements Compliance Electromagnetic environment – guidelines
HF emissions in accordance with
CISPR 11
Group 1 The iLED 7 surgical light system uses HF
energy only for its internal functioning.
Therefore, its HF emissions levels are very
low and it is improbable that neighboring
devices would be affected by
interference.
HF emissions in accordance with
CISPR 11
Class A The iLED 7 surgical light system is intended
for operation in facilities other than private
homes, provided that these facilities are
directly connected to a public power
supply network that also supplies buildings
used for residential purposes.
Harmonic emissions as per
EN/IEC 61000-3-2
Class A
Voltage fluctuations / flickers in
accordance with EN/IEC 61000-3-3
fulfilled
HF emissions in accordance with
CISPR 11
Class B The Mobile Control 7,9 is suitable for use in
all establishments including private homes
and those directly connected to a public
power supply network that also supplies
buildings used for residential purposes.
Harmonic emissions as per
EN/IEC 61000-3-2
Class B
Voltage fluctuations / flickers in
accordance with EN/IEC 61000-3-3
fulfilled
Guidelines and manufacturer's declaration – electromagnetic immunity
The iLED 7 surgical light system/Mobile Control 7,9/Wall Control Panel is intended for use in the
electromagnetic environment indicated below. The customer or user of the iLED 7 surgical light
system/Mobile Control 7,9/Wall Control Panel must ensure that it is operated in such an environment.
Immunity testing EN/IEC 60601-1 test
level
Compliance level Electromagnetic environment –
guidelines
Static electricity
discharge (ESD)
according to
EN/IEC 61000-4-2
± 8 kV contact
discharge
± 15 kV air discharge
± 8 kV contact
discharge
± 15 kV air discharge
The floor should be made of
wood or concrete, or should be
covered with ceramic tiles. If a
floor is covered with synthetic
material, the relative humidity
must be at least 30%.