3-5
3.4.3 Network Verification Process
This part is completed through two ways: First the hospital completes items requiring self-check of the hospital’s IT
Dept, as indicated in the Network Verification Table. Then customer service personnel or authorized party confirms items
on site and finally completes the Network Acceptance Table. If any item is found incompliant during network Verification
test, adjustment should be made before the patient monitors installation.
When in test, the Wi-Fi network SSID broadcast needs to be enabled to ensure that the Wi-Fi SSID can be scanned.
Table 3-2 Wireless Network Acceptance Table
No. Item Content of Requirements Verification Method Check Results
Wireless coverage requirements
1 Received
signal
strength
(RSSI)
≥-65 dBm
RSSI is the value displayed on the
patient monitor
Service person performs the test
by using network survey tool.
Make sure that all expected
coverage areas such as ward,
corridor, toilet, stairs, and elevator
are tested.
2 Co-
channel
interferen
ce
≤-20dB
Service person performs the test
by using network survey tool.
Make sure that all expected
coverage areas such as ward,
corridor, toilet, stairs, and elevator
are tested.
3 Ping
delay
The mean delay of PC or cell phone
with normal Wi-Fi module is smaller
than 100ms and The packet lost
rate shall be less than 1%.
Service person performs the test:
1. Connect PC or cell phone with
normal Wi-Fi module to A P.
2. Connect another PC to the LAN
port where the central
monitoring system is connected
to.
3. Run the command” “ping –t –l
32 –w 1000 IPaddress-of -
cellphone” for 10 minutes.
4. Run“ ctrl+c”.
Requirements of AP capability