4-1
4 Testing and Maintenance
4.1 Introduction
To ensure the patient monitor always functions properly, qualified service personnel should perform regular inspection,
maintenance and test. This chapter provides a checklist of the testing procedures for the patient monitor with
recommended test equipment and frequency. The service personnel should perform the testing and maintenance
procedures as required and use appropriate test equipment.
The testing procedures provided in this chapter are intended to verify that the patient monitor meets the performance
specifications. If the patient monitor or a module fails to perform as specified in any test, repairs or replacement must be
done to correct the problem. If the problem persists, contact our Customer Service Department.
CAUTION
All tests should be performed by qualified service personnel only.
Care should be taken when changing the settings in Maintenance and Configuration menus to avoid loss
of data.
Service personnel should possess a working knowledge of the test tools and make sure that test
equipment and cables are applicable.
4.1.1 Test Equipment
See the following sections.
4.1.2 Test Report
Upon completion of the tests, the table of preventative maintenance test reports and the table of maintenance test
reports in this chapter should be kept properly.
4.1.3 Preventative Maintenance
The following sections provide a list of recommended preventative maintenance procedures. It is recommended to
maintain the patient monitor at least once every two years (and once a year for CO
2
and AG modules). (See the following
sections for detailed test procedures and contents)
4.1.4 Recommended Frequency
Check/Maintenance Item Frequency
Preventative Maintenance Tests
Visual inspection When first installed or reinstalled.
NIBP tests
Pressure check
1. If the user suspects that the measurement is incorrect.
2. Following any repair or replacement of relevant module.
3. For NIBP module, at least once every two years; for CO
2
and AG
modules, once a year.
4. AG leakage test should be performed before AG measurement.
Leakage test
Sidestream and
Microstream CO
2
tests
Leakage test
Performance test
Calibration