EasyManua.ls Logo

BeneVision N17 - Whole Unit Disassembly; Basic Disassembly

Default Icon
146 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7-2
7.3 Whole Unit Disassembly
7.3.1 Basic Disassembly
NOTE
Be sure to disassemble the base first before proceeding with other parts.
1. Remove the power plug anti-pull hook from the rear case of the monitor.
2. Use a tweezer to pry up the four screw covers at the four corners of the rear case and loosen and remove the four
M3×8 screws under them.
(Screw cover) (Screw)

Table of Contents