6-1
6 Upgrade
6.1 Overview
This monitor supports upgrade of the monitoring parameter function modules, upgrade of the functional assemblies,
and network upgrade of software.
For function upgrade involving disassembly of the monitor, eliminate static electricity before the
disassembly. When removing some parts with the electrostatic sensitive mark, wear protective devices
such as electrostatic ring or anti-electrostatic gloves, lest the parts would be damaged.
Properly connect and route the cables and wires when reassembling the equipment to avoid pinched
hoses and electrical short circuits.
Use specified screws to reassemble the equipment. If the incorrect screws are forcefully tightened, the
equipment may be damaged and the screws or part may fall off during use, causing unpredictable
equipment damage or human injury.
Be sure to follow the correct sequence when disassembling the monitor.
Before removing assemblies, make sure that all the connection lines have been unplugged. During
removal, note to avoid breaking the connection line by pulling or damaging the connector.
Place the removed screws and other parts separately by category so that they can be used in the re-
installation. Do not drop, contaminate or lose them.
6.2 Upgrade of Parameter Function Modules
You can upgrade the following parameter modules:
Monitoring Parameter
Module
PN Name and Specification Remarks
MPM module 115-038696-
00
MPM-1 Module (MR Spo
2
/3/5 lead/IBP)
115-038697-
00
MPM-3 Module (NC Spo
2
/3/5 lead/IBP)
115-044666-
00
MPM-7 Module (MR Spo
2
/3/5lead/No
IBP)
115-044667-
00
MPM-9 Module (NC Spo
2
/3/5lead/No
IBP)
115-038698-
00
MPM-13 (MR SpO
2
/12lead/IBP/Analog)