6-13
6.5 Upgrading CAA license function
To upgrade the monitor to have CAA functions, you need to buy the licenses. Provide the following information when
you order the licenses:
Order number (contents CAA license function requirement )
Product Model (for example, BeneVision N17/M17)
Monitor serial number (for example, F2-6C000031)
MID number/MAC address (for example: 000F140839AE). To get the MID number, select : Main Menu → License
(in the System column).
Customer's email (for receiving license key file)
Refer to the following instructions to upgrade the licenses
1. Unzip the PMLS.zip file. Then you would get a folder named PMLS.
2. Copy the folder to a USB disk.
3. Plug the USB disk into the BeneVision N series patient monitor.
4. In the interface of BeneVision monitor, choose Main Menu→License (in the System column)→External, and then
click the Install button.
The U disk should be FAT32 format.
The PMLS folder should be in the root directory.
On the root folder should be the PMLS folder and inside the PMLS folder should be the folder named
same as MID number, and in there should be the license files.