4-23
The message Calibration Completed! is displayed after a successful calibration. Otherwise, a corresponding message
will be displayed.
C.O. Test
Method 1:
Tools required:
Medsim300B Patient simulator or Prosim 8, or equivalent equipment
C.O. adapter box (for 300B)
PiCCO C.O. test cable (PN: 040-001301-00)
1. Connect the patient simulator and the C.O. module using a C.O. trunk cable and a C.O. adapter box.
2. Set the blood temperature (BT) to 37ºC on the patient simulator and check the temperature value is 37±0.1ºC.
3. Select CCO parameter interface to access the C.O. Measure interface. Then, select Setup to check whether the value
of Catheter Type is PV2015L20.
4. Turn the injectate temperature (TI) knob on the C.O. adapter box to set the TI to 20±1ºC for the patient simulator
and C.O. to 5 L/min.
5. In the C.O. Measurement window, select Start to start C.O. measurement. As soon as the prompt Inject XXml is
displayed, adjust TI to 4±1ºC, and then quickly back to 20±1ºC. Simultaneously press the button on the simulator
that corresponds to 5 L/min. The whole procedure shall be finished in 10 seconds.
6. Verify that the C.O. value displayed on the monitor is correct.
Method 2:
Tools required:
Medsim300B Patient simulator or Prosim 8, or equivalent equipment
C.O. adapter box (for 300B)
PiCCO C.O. test cable (PN: 040-001301-00)
1. Connect the patient simulator and the C.O. module using a C.O. trunk cable and a C.O. adapter box.
2. Set the blood temperature (BT) to 37ºC on the patient simulator and check the temperature value is 37±0.1ºC.
3. Select CCO parameter interface to access the C.O. Measure interface. Then, select Setup to check whether the value
of Catheter Type is PV2015L20.
4. Turn the injectate temperature (TI) knob on the C.O. adapter box to check whether the value of the TI is from 0 to
30ºC.
4.4.16 NMT Tests
Performance Test
Method 1:
Tools required:
Resistance box
Multimeter