EasyManua.ls Logo

BeneVision N17 - NIBP Tests; Temp Test; IBP Tests

Default Icon
146 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4-13
A simulator cannot be used to assess the accuracy of a pulse oximeter monitor or a SpO
2
sensor. Instead,
it can only verify that whether the monitor is functional. The accuracy of a pulse oximeter monitor or a
SpO
2
sensor needs to be verified by clinical data.
4.4.4 NIBP Tests
See section 4.2.24.2.2NIBP Tests.
4.4.5 Temp Test
Tools required:
Resistance box (with accuracy above 0.1 Ω)
1. Connect the two pins of any Temp connector of a module to the two ends of the resistance box using two wires.
2. Set the resistance box to 1354.9 Ω (corresponding temperature is 37ºC).
3. Verify each Temp channel of the monitor and make sure that the displayed value is within 37±0.1ºC.
4.4.6 IBP Tests
Performance Test
Tools required:
Medsim300B patient simulator, MPS450, or other equivalent equipment
Dedicated IBP adapter cable (300B, P/N 00-002199-00) (use P/N 00-002198-00, if the simulator is MPS450)
1. Connect the patient simulator to the monitor's IBP connector.
2. Set the patient simulator output to the IBP channel to 0 mmHg.
3. Press the Zero key on the module to make a zero calibration.
4. Set static pressure to 200 mmHg on the patient simulator.
5. The displayed value should be within 200±2 mmHg.
6. If the value is outside of these tolerances, calibrate the IBP module. If the IBP module was calibrated with a
dedicated reusable IBP sensor, check the calibration together with this IBP sensor.
7. Make the patient simulator outputs 120/80 mmHg ART signals and 120/0 mmHg LV signals respectively to each IBP
channel and check that the IBP wave is displayed correctly.
Pressure Calibration
Method 1:
Tools required:
Medsim300B patient simulator, MPS450, or other equivalent equipment
Dedicated IBP adapter cable (300B, P/N 00-002199-00) (use P/N 00-002198-00, if the simulator is MPS450)
1. Connect the patient simulator to the monitor's IBP connector.

Table of Contents

Related product manuals