LB 134 9. Basis of Calculation
A
RR
BgrdSource
Detection
BgrdSampleCALSurface
RRNA
W
RR
A
Detection
BgrdSample
Surface
(4)
The ISO calibration is performed according to the International
Standard ISO 7503-1 [1] relative to the surface emission rate
q
2
of the reference source and the sensitive detector surface
W, as a reference surface. In accordance with ISO 7503-1, the
equations 5-8 also take into account the emission efficiency
Source
.
2
q
RR
BgrdSource
Detection
sourceDetection
CAL
W
N
1
BgrdSampleCALSurface
RRNA
sourceDetection
BgrdSample
Surface
W
RR
A
(8) (Bq/cm²)
In this formula the area W is explicitly included, therefore use
for the emission rate q
2π
the total number of betas per second.
If the surface emission rate is not available, as in this case, we
use
The emitter efficiency
source
of alpha emitters must be assumed
to be 0.25, that of beta emitters 0.50 [1]. For some beta emitters
with low beta energies we also have to use 0.25. If the sensitive
detector area is larger than 300 cm
2
, 300 cm
2
is used for W.
[1] DIN ISO 7503-1:1990-07: Evaluation of surface contamina-
tion; beta-emitters (maximum beta energy greater than
0,15 MeV) and alpha-emitters; identical with ISO 7503-
1:1988