With LB 1342 Alpha-Beta contamination probe, Counter/Timer, 10s measuring time, 3 cycles
Explanation:
LB1342 (1) (LB Number of probe)
22.02.2015, 16:36:03 (Date and Time)
NE 0.0 cps 0.0 cps 600 s (Background and BG Time)
Netto 1.000 100.0 cps 0.6300 (C-Factor and alarm threshold)
Netto 1.000 100.0 cps 0.6300 (C-Factor and alarm threshold)
22.02.2015 16:36:13 10 s a:Netto 0.10000 cps 100.0 % 0.0 cts 0.0 h
LB1342 ZT bg:Netto 5.451 cps 13.5 % 0.0 cts 0.0 h
22.02.2015 4:36:23 10 s a:Netto 0.10000 cps 100.0 % 0.0 cts 0.0 h
LB1342 ZT bg:Netto 8.351 cps 10.9 % 0.0 cts 0.0 h
22.02.2015 4:36:33 10 s a:Netto 0.0 cps 100.0 % 0.0 cts 0.0 h
LB1342 ZT bg:Netto 8.067 cps 11.1 % 0.0 cts 0.0 h
22.02.2015 4:36:33 30 s a:Netto 0.06667 cps 70.7 % 0.0 cts 0.0 h
LB1342 ZT M bg:Netto 7.291 cps 6.7 % 0.0 cts 0.0 h
Explanation:
Date Time M.Time s Nuclidename1 M.value1 unit1 accuracy1 % Intvalue1 IntUnit1 IntTime h
LB-Number Mode Nuclidename2 M.value2 unit2 accuracy2 % Intvalue2 IntUnit2 IntTime h
The last two lines belong to the mean value over the three cycles.
In a single-channel probe, e.g. doserate probe, the measured values are missing in the second row.