d) --FIFO EMPTY--
Data memory empty or write pointer on 1
st
position
e) --FIFO TOP--
Data memory read pointer has reached the write pointer.
Separator between response data is the colon.
Example of frame with F2C protocol:
Command: Get current or oldest FIFO element for all probes
Command to LB 134:
*0001900101010100**CR
Response from LB 134:
*900100010201010022.02.201517:50:03:0037:0002:002024040000:0030:Netto:0:cps:0:cts:DosisL1:0.106536
:µSv/h:0.000387:µSv:0.003056:0:0:0:0.07:0000:LB1346*70*
Meaning of parameters and measured value of the response:
*FDV TDV Probe code Date Time : No.of FIFOelement : Zyclus number : 12xAsciihexStatus : M.Time : Nuclide1
: M.Value1 : unit1 : Intvalue1 : Intunit1 : Nuclide2 : M.value2 : unit2 : Intvalue2 : Intunit2 : IntTime : Rawdata1 :
Background1 : Rawdata2 : Background2 :No.Zycles : LB1346 * Checksume *
Example for transferred data during automatic data transfer to the PC:
With LB 1342 Alpha-Beta contamination probe, ratemeter, 10s measuring time, 0 cycles (endless)
22.02.2015 16:16:52 10 s a:Netto 1.16E-03 cps >100 % 0.0 cts 0.0 h
LB1342 RM bg:Netto 8.438 cps 2.2 % 0.0 cts 0.0 h
22.02.2015 16:17:02 10 s a:Netto 1.06E-03 cps >100 % 0.0 cts 0.0 h
LB1342 RM bg:Netto 8.356 cps 2.2 % 0.0 cts 0.0 h
22.02.2015 16:17:12 10 s a:Netto 9.79E-04 cps >100 % 0.0 cts 0.0 h
LB1342 RM bg:Netto 8.357 cps 2.2 % 0.0 cts 0.0 h
Explanation:
Date Time M.Time s Nuclidename1 M.value1 unit1 accuracy1 % Intvalue1 IntUnit1 IntTime h
LB-Number Mode Nuclidename2 M.value2 unit2 accuracy2 % Intvalue2 IntUnit2 IntTime h