EasyManua.ls Logo

Berthold LB 134 - 5.4 Exceeding of Limit Values

Berthold LB 134
157 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
LB 134 5. Contamination Measurements
40
Figure 4.6: Background menu for contamination probe
Set the desired parameters for the background measure-
ment.
Background: Shows the currently stored -
background. It may come from a back-
ground measurement or it can be entered
manually. The stored background is al-
ways subtracted from each measure-
ment.
Background: Shows the currently stored
background. It may come from a
background measurement or it can be
entered manually. The stored back-
ground is always subtracted from each
measurement.
Meas. Time: Enter the desired measuring time. Enter
at least 60 seconds, better 600 seconds
in order to get accurate results.
Autostart: When ON is set (X mark), a background
measurement is carried out automatically
whenever the device is turned on.
BG Meas. Time: If a background measurement was car-
ried out, the elapsed background meas-
urement time is entered here after sav-
ing. On the other hand, if you have en-
tered the background by hand, you have
to enter the measuring time associated
with the background, so that the accura-
cy calculation is correct for a contamina-
tion measurement.
Start the background measurement by moving the cursor
() to Measure background and press the Enter key.
Then the background measurement starts.
Figure 4.7: Background measurement
The measured values and the elapsed measuring time are
displayed continuously. At the end of the measuring time,

Table of Contents

Related product manuals