KBF P + KBF LQC + KBWF (E6) 11/2016 page 169/189
Climatic data (with humidity) for KBWF
Temperature range without light cassettes
°C / °F
Temperature range with light cassettes, with illumination
°C / °F
+10 to +60 /
+20 to +60 /
Temperature fluctuation with illumination
Temperature uniformity (variation) with illumination
Humidity range without light cassettes
Humidity range with light cassettes, with illumination
Humidity fluctuation with illumination
Illumination data per light cassette
ICH compliant illumination for photo stability testing
Daylight fluorescent tubes
Fluora
®
growth lamps
Arabidopsis lamps
Electrical data (model versions KBFP-230V, KBFLQC-230V, KBWF-230V)
IP system of protection acc. to EN 60529
Nominal voltage (+/-10%) at 50 Hz power frequency
Installation category acc. to IEC 61010-1
Pollution degree acc. to IEC 61010-1
Over-current release category B, 2 poles
Different electrical data for UL chambers constructed for the USA and Canada
(model versions KBFP240UL-240V, KBFP720UL-240V, KBFLQC240UL-240V, KBFLQC720UL-240V)
Nominal voltage (+/-10%)
Environment-specific data
Energy consumption at 40 °C / 104 °F and 75 % r.H.
Filling weight of refrigerant R 134A (GWP 1300)
Illumination data: Average value, measured at +22 °C +/- 3 °C / 71.6 °F +/- 5.4 °F with a spherical sen-
sor (+/-10%) by 12 cm / 4.7 in below the light cassette. The values given in W/m
2
refer to global radiation.
All technical data is specified for unloaded chambers with standard equipment at an ambient temperature
of +22 °C +/- 3°C / 71.6 °F +/- 5.4 °F and a power supply voltage fluctuation of +/-10%. Technical data is
determined in accordance to BINDER factory standard Part 2:2015 and DIN 12880:2007.
All indications are average values, typical for chambers produced in series. We reserve the right to
change technical specifications at any time.
If the chamber is fully loaded, the specified heating up and cooling down times may vary ac-
cording to the load.
Bringing a source of humidity into the inner chamber will affect the minimum humidity specifica-
tion and may affect the humidity accuracy.