List Mode 54
Step Parameters
Voltage Sets Vset value.
When setting Vset the list will not limit the voltage range based on mode.
Therefore, Vmax will be 96.960 for model 9140 and 180.800 for model 9141.
To prevent a protection trip, verify the Vset and Iset values in the list is within
the range of the congured operation mode before running the list.
Current Sets Iset value.
When setting Iset the list will not limit the current range based on mode.
Therefore, Imax will be 24.240 for model 9140 and 12.120 for the 9141.
To prevent a protection trip, verify the Vset and Iset values in the list is within
the range of the congured operation mode before running the list.
BOST Press the knob or to enable/disable a trigger-out signal at the beginning of the step (BOST).
EOST Press the knob or to enable/disable a trigger-out signal at the beginning of the step (EOST).
Dwell Sets the dwell time of the selected step.