82
Z31980_00_03
Configuration
Hex File
5. Reset the Master CPU (in this example the VTU) by pressing the “Hardware Reset”
Button.
Shim Control: Main Board: main -> shim -> base function
This procedure takes about
30 seconds and is finished
when the programming menu
reappears.
4. Repeat this procedure (Upload
& then flash) for the Hex file.
This procedure takes a few min-
utes and is finished when the fol-
lowing message appears:
6. Check that the correct version
has been successfully updated.