94
Z31980_00_03
Configuration
Values”
Get Dac values:
This function fetches the active DAC-Values from the Shim Control Board controller to
the Web page.
Set DAC-Values to coils:
This function sets the active DAC-Values in the Shim Control Board controller to all
coils.
Apply:
This button will apply the function you have chosen with the checkboxes.
Coil 2
On this page you can see the 8 coils which are controlled by the second Shim Control
Board. The resolution of the DAC which set the current to the coils is 16 bits.